■ Display mode

High Definition 8.4 TFT color LCD display


Ventilation mode

IPPV, SIPPV, SIMV, PEEP, MANU, SIGH


ventilation parameters

Tidal volume of 50 ~ 1500ml
Frequency of 2 to 99 times / min
Absorption ratio of 2:1 ~ 1:8 Call SIMV frequency of 2 to 20 times / min Inspiratory trigger pressure -1. 0 ~ 2.0 kPa PEEP 0 ~ 2.0 kPa
1.0 ~ 6.0 kPa pressure control Sigh inspiratory time 1.5 times Inspiratory Ping Taiwan 0 ~ 50% Oxygen concentration of 45% to 100%


Ventilation monitoring

Tidal volume, ventilation, IPPV frequency, SIMV frequency, total respiratory rate, ratio, peak airway pressure, mean airway pressure, pressure - time waveform, flow rate - the time wave, positive end expiratory pressure, inspiratory trigger pressure , inspiratory Ping Taiwan, oxygen concentration


Security Alarm System

Maximum oxygen concentration setting range from 50% to 100%
Limit setting range of 15% to 50%
Airway pressure alarm limit setting range of 1.0 ~ 6.0 kPa
Lower limit range of 0.4 ~ 2.0 kPa
Minute ventilation alarm limit setting range 3.0 ~ 30 L / min
Limit setting range 1.0 ~ 10 L / min
High pressure continued to alarm when the alarm has been higher than 2.5 kPa
Suffocation alarm input 15s without tidal volume, the sound, light alarm
Power failure alarm

Power Supply AC 220V ± 10% 50Hz ± 1Hz UPS power supply, battery

arm